Minimum Test Under Life-Threatening Conditions
Minimum Test Under Life-Threatening Conditions
When situations do arise in anesthesia for trauma or emergency cesarean section where there is neither time nor opportunity to fully check the anesthesia gas machine. The following checklist is suggested for these situations. It requires little if any additional time, and can add greatly to safety, and hence, peace of mind.
- High pressure test of the breathing circuit: Ensures there are no leaks distal to common gas outlet
- Check patient suction
- Observe and/or palpate breathing bag during Preoxygenation. This ensures
- Adequate flow of oxygen
- Good mask fit (very important)
- The patient is breathing
- The Bag/Vent switch is on "Bag" not "Vent"
Last modified: Wednesday, 16 November 2016, 11:57 AM